9363133 rev. C 0704
16
4.1 Troubleshooting
Furnace does not turn on
• Disconnect power cord from furnace, check fuses at inlet socket on back panel (230V).
Check F1 on PCB
• Check J-16 on the control PCB
• Exchange control PCB (chapter 7.1.1)
LCD display has no content
• Check contrast adjustment - turn adjustment wheel after furnace is turned on (first
display screen that appears).
• Check the -15V on the control PCB (U24 : 1)
• Exchange display PCB (chapter 7.1.3)
• Exchange control PCB (chapter 7.1.1)
Membrane keys without function
• Exchange membrane (chapter 7.1.2)
• Exchange control PCB (chapter 7.1.1)
Adjustment wheel without function
• Exchange encoder PCB (chapter 7.1.4)
• Exchange control PCB (chapter 7.1.1)
Piston moves too slow or too fast (Cerampress) (units built prior to date code 0232)
• Check air pressure (4.5 bar)
• Check air flow restriction adjustment (See Fig. 1)
• Disconnect then reconnect power cord from furnace
• Observe piston down/up movement
• If down speed is too fast, turn bottom screw, located on the pressing valve (Sol 3), slightly
clockwise
• If up speed too fast, turn top screw, located on the pressing valve (Sol 3), slightly clockwise
Pressing Valve (Sol 3)
Fig. 1
Adjustment for
up movement speed
Adjustment for
down movement speed