4.1 Troubleshooting
Furnace does not turn on
Disconnect power cord from furnace, check fuses at inlet socket on back panel (230V).
Check F1 on PCB
Check J-16 on the control PCB
Exchange control PCB (chapter 7.1.1)
LCD display has no content
Check contrast adjustment - turn adjustment wheel after furnace is turned on (first
display screen that appears).
Check the -15V on the control PCB (U24 : 1)
Exchange display PCB (chapter 7.1.3)
Exchange control PCB (chapter 7.1.1)
Membrane keys without function
Exchange membrane (chapter 7.1.2)
Exchange control PCB (chapter 7.1.1)
Adjustment wheel without function
Exchange encoder PCB (chapter 7.1.4)
Exchange control PCB (chapter 7.1.1)
Piston moves too slow or too fast (Cerampress) (units built prior to date code 0232)
Check air pressure (4.5 bar)
Check air flow restriction adjustment (See Fig. 1)
Disconnect then reconnect power cord from furnace
Observe piston down/up movement
If down speed is too fast, turn bottom screw, located on the pressing valve (Sol 3), slightly
clockwise
If up speed too fast, turn top screw, located on the pressing valve (Sol 3), slightly clockwise
Pressing Valve (Sol 3)
Fig. 1
Adjustment for
up movement speed
Adjustment for
down movement speed